| Technical Specifications | |
| Product Category | Electronic Sphygmomanometers |
| Product description | Automatic Upper Arm Blood Pressure Monitor |
| Model (code) | HBP-1320 (HBP-1320-E) |
| Measurement Parameter | NIBP, PR |
| Dimension | Main unit: 123 x 201 x 99 (mm) 4.84 x 7.91 x 3.90 (inch) (W x H x D) AC adapter: 46 x 66 x 66.2 (mm) 1.81 x 2.60 x 2.61 (inch) (W x H x D) Battery: 54 x 43.5 x 15.4 (mm) 2.13 x 1.71 x 0.61 (inch) (W x H x D) |
| Weight | Main unit: Approx. 0.52 kg (not including accessories and options) AC adapter: Approx. 114 g Battery: Approx. 0.1 kg |
| Display | 7 segment LCD |
| Protection Class | Class II (AC Adapter) Internal powered equipment (when operating with battery only) |
| Degree of Protection | Type BF (Applied part): Cuff |
| Operation Mode | Continuous operation |
| MDR Classification | Class II a |
| Power supply | |
| AC adapter | Input voltage range: AC 100 V to 240 V Frequency: 50/60 Hz Output voltage range: DC 6 V ±5% Rated output current: 1.6 A |
| Rechargeable battery | Type: 3.6 V, 1900 mAh Number of operation cycles when fully charged: 300 Measurement conditions • New battery fully charged • Ambient temperature of 23°C (73.4°F) • Using M-size cuff • SYS 120 / DIA 80 / PR 60 (Inflation setting: AUTO) • One 5-minute cycle consisting of “cuff measurement time + wait time” |
| Environmental Conditions | |
| Operational temperature and humidity | Temperature range: 5 to 40°C (41 to 104°F) Humidity range: 15 to 85%RH (not condensed) Atmospheric pressure: 700 to 1060hPa |
| Storage and transportation | Temperature range: -20 to 60°C (-4 to 140°F) Humidity range: 10 to 95%RH (not condensed) Atmospheric pressure: 500 to 1060hPa |
| Non-Invasive Blood Pressure (NIBP) | |
| Measurement technology | Oscillometric |
| Measurement method | Dynamic Linear Deflation method |
| Pressure display range | 0 to 300 mmHg |
| Pressure display accuracy | Within ±3 mmHg |
| NIBP measurement range | SYS 60 to 250 mmHg DIA 40 to 200 mmHg Pulse rate 40 to 200 /min |
| NIBP accuracy | Maximum mean error within ±5mmHg Maximum standard deviation within 8mmHg |
| Pulse rate accuracy | Within ±5 % of reading |