| Technical Specifications: HBP-1120 | |
| Product category | Electronic Sphygmomanometers |
| Product description | Automatic Upper Arm Blood Pressure Monitor |
| Model (code) | HBP-1120 (HBP-1120-E) |
| Measurement Parameter | NIBP, PR |
| Dimension | Main unit: 130 × 175 × 120 (mm) 5.12 × 6.89 × 4.72 (inch) (W×H×D) AC adapter: 64.5 × 21 × 51 (mm) 2.53 × 0.83 × 2.01 (inch) (W×H×D) |
| Weight | Main unit: Approx. 510 g (not including accessories) AC adapter: Approx. 48 g |
| Display | 7 segment LCD |
| Protection Class | Class II (AC adapter) Internal powered equipment (when operating with battery only) |
| Degree of Protection | Type BF (Applied part): Cuff |
| Operation Mode | Continuous operation |
| MDR Classification | Class II a |
| Power supply | |
| AC adapter | Input voltage range: AC 100 V to 240 V Frequency: 50/60 Hz Output voltage range: DC 6 V ±0.5 V Rated Output Current: 0.7 A |
| Dry cell battery | Type: AA batteries, x4 Approx. 250 measurements • Measurement conditions – New batteries (AA high-performance manganese) – Ambient temperature of 23°C (73.4°F) – Using M-size cuff – SYS120 / DIA80 / PR60 – One 5-minute cycle consisting of “cuff measurement time + wait time” |
| Environmental Conditions | |
| Operating Conditions | Temperature range: 5 to 40°C (41 to 104°F) Humidity range: 15 to 85%RH (not condensed) Atmospheric pressure: 700 to 1060hPa |
| Storage and transportation | Temperature range: -20 to 60°C (-4 to 140°F) Humidity range: 10 to 95%RH (not condensed) Atmospheric pressure: 500 to 1060hPa |
| Non-Invasive Blood Pressure (NIBP) | |
| Measurement technology | Oscillometric |
| Measurement method | Dynamic Linear Deflation method |
| Pressure display range | 0 to 300 mmHg |
| Pressure display accuracy | Within ±3 mmHg |
| NIBP measurement range | SYS 60 to 250 mmHg DIA 40 to 200 mmHg PULSE 40 to 200 /min |
| NIBP accuracy | Maximum mean error within ±5 mmHg Maximum standard deviation within 8 mmHg |
| Pulse rate accuracy | Within ±5 % of reading |